Journal: Scandinavian journal of clinical and laboratory investigation
Article Title: Endogenous anti-streptavidin antibodies causing erroneous laboratory results more common than anticipated.
doi: 10.1080/00365513.2020.1858493
Figure Lengend Snippet: Figure 4. Comparison of results from the Roche method and alternative methods in the 42-sample set sent to Roche Diagnostics for further investigation and in the 51- sample set not investigated for the source of interference. Concentrations above and below the detection limit were given the detection limit values. Values obtained by the Roche method were adjusted for method dependent differences according to the equations listed in Table 1. Median, 25 and 75 percentiles and 95% confidence intervals are shown. (A) TSH. 42-sample set: 35 TSH samples measured with Roche and DELFIA method, median TSH Roche concentration was 1.03 mIU/L (range 0.07–64.20), median TSH DELFIA concentration was 1.86 mIU/L (range <0.03 to >100). 51-sample set: 44 TSH samples measured with both methods, median Roche concentration was 1.12 (range 0.08–78.46), median DELFIA concentration was 1.86 mIU/L (range 0.32–10.70). Y-axis uses logarithmic scale. (B) FT4. 42-sample-set: 37 FT4 samples measured with Roche and DELFIA method, median FT4 Roche concentration was 28.8pmol/L (range 5.7 to >78.8), median FT4 DELFIA concentration was 14.1pmol/L (range 6.2–26.0). 51- sample set: 47 FT4 measured with both methods, median FT4 Roche concentration was 28.0pmol/L (range 14.9 to >78.8), median FT4 DELFIA concentration was 14.6pmol/L (range 10.6–33.4). (C) FT3. 42-sample set: 25 FT3 samples measured with Roche and alternative method, median FT3 Roche concentration was 10.5pmol/L (range <1.9 to >39.2), median FT3 concentration with alternative methods was 6.3pmol/L (range 3.9–9.6). 51-sample set: 27 FT3 measured with two methods, median FT3 Roche concentration was 8.4pmol/L (range 5.4–14.9), median FT3 with alternative methods was 5.9pmol/L (range 2.5–7.5).
Article Snippet: Dissociation enhanced lanthanide fluorescence immunoassay (DELFIA, Perkin Elmer) was used for the determination of serum TSH (reference range, RR: 0.5–3.6 mIU/L), FT4 (RR: 8.0–21.0 pmol/L) and until February 2016 for determination of FT3 (RR: 3.5–8.3 pmol/L).
Techniques: Comparison, Concentration Assay